[Front View]
[Autosampler]
[Operator Panel]
[Column Oven]
[Pumping Unit & UV Detector (bult-in option)]
[De-gas / Purge Unit]
High Reliability
When the system is used with semi-micro SEC columns, featuring calibration curves with superior linearity*, and the standard polystyrene kit PStQuick, which ensures minimization of preparation errors, highly reliable data with minimized variation between systems may be obtained.
(*Compared with Tosoh's conventional products)
High Stability
The column oven has a double structure and a double temperature-control mechanism.
A double-path, double-flow refractive index detector (RI detector) is used for temperature control of the main units, including tubing. A stable baseline is obtained by suppressing fluctuation of the refractive index in the solvent.
High Reproducibility
The pumping unit features a newly designed structure and control system. Temperature control of the entire pumping unit gives high reproducibility which is not affected by varying solvent conditions or environmental temperature fluctuations.
Fast, Efficient, Low Solvent Consumption
Combined use with TSKgel semi-micro SEC columns (SuperMultiporeHZ, SuperHZ, SuperH, and SuperAW) allows fast, efficient separation and SEC analysis and reduces solvent consumption.
Easy Operation
Using the GPC Workstation EcoSEC-WS, the system can easily be operated through warm-up, analysis, and shut-down. The main unit may also be independently controlled using the operator panel.
The automatic purge function eliminates solvent replacement
and other time-consuming manual operations.
Global Standard
Compliant with China RoHS.
UL61010-1 and CE mark certification.
Also compliant with FDA 21 CFR PART 11*.
(*Authentication using user ID and password, logout when finished, software validation, etc.)
UV-8320
Part # 0021792
No extra installation space is necessary for this built-in option.
As it can be used in the pump oven, the detector is highly stable.
Specifications | Description |
Method | Dual beam, single flow cell |
Wavelength | 195 – 350 nm |
Flow Cell | Volume | 2 µL |
| Maximum pressure | 3 MPa |
| Noise | 2.5 × 10-5 ABU |
| | (254 nm, cell containing air, response: 1.0 s) |
| Drift | 3.0 ×10-4 ABU/h |
| | (254 nm, cell containing air, response: 1.0 s) |
Safety Features | Leak sensor, lamp lighting monitor |
| Pan for liquid leakage in the installation area |
Weight | 7 kg |
Column Switching Valve Unit
Part # 0021793
Two series of analytical columns can be connected to this electric-powered two-series four-way switching valve. The valve allows the easy switching of high-and low-molecular columns | |
* When the following organic solvents are used as eluent, alternative rotor seal, P/N 0023288 should be used. DMF (dimethylformamide), DMAc (dimethylacetamide), NMP (n-methylpyrrolidone) |
Sample Cups
Part # | Description | Quantity |
0022006 | Glass cup for use with septum 2.0 mL | 1000 |
0022007 | Cap for use with septum | 1000 |
0022009 | Septum packing | 1000 |
0022008 | Glass cup for use with septum 2.0 mL | 500 |
| (Cap + packing) | |
0022115 | Glass cup - Transparent 2.0 mL | 1000 |
0022116 | Glass cup - Brown 2.0 mL | 1000 |
0023213 | Glass cup - Transparent 1.1 mL | 1000 |
0016330 | Cap | 1000 |
0013718 | Aluminum sheet | 100 |
0013717 | Packing | 100 |
0016618 | Glass cup - Transparent 2.0 mL | 500 |
| Cap + aluminum sheet + packing) | |
HLC-8321GPC Specifications
Specifications | Description |
Degas Unit |
Method | Vacuum Degassing method |
| Vacuum selection from two levels |
Volume | Solvent: 20 mL x 2 series (selectable) |
| Wash: 10 mL |
Pump Oven Unit |
Method | Hot-air agitation, PID control |
| Room temperature +10 to 50 ºC ( 1 ºC steps) |
| Temperature setting accuracy: ± 0.5 ºC |
| Temperature control precision: ± 0.1 ºC |
Safety Features | Temperature sensor for oven temperature monitor. |
| Temperature control stops if abnormality is detected |
| Gas sensor for stopping temperature control |
| Thermal fuse (70 ºC) for temperature control circuit block |
| Door status monitor: Temperature control stops when door is opened |
| Pan for liquid leakage |
Pump |
Liquid supply mode | Parallel liquid supply |
Flow Rate | 10 to 2000 µL/min ( 1µ/min steps ) |
| (Sample flow rate only: Reference flow rate is set by ratio |
| [1/2, 1/3, 1/4] to the sample flow rate) |
Accuracy | ± 2 % or less OR ± 5 µL/min or less, whichever is greater |
| (when H2O supply pressure is 1 MPa or more, 1200 µL/min or less) |
Precision | ± 2 % or less OR ± 1 µL/min or less, whichever is greater |
| (when H2O supply pressure is 1 MPa or more, 1200 µL/min or less) |
Max Pressure | 0 to 25 MPa (0.1 MPa steps) |
Safety features | Liquid supply stops if pressure rises above the upper limit or |
| drops below lower limit |
| Plunger drive count monitoring |
| Pan for liquid leakage |
Autosampler |
Method | Syringe measurement |
Standard loop | 100 µL |
Injection volume | 1 to 1500 µL (1 µL steps) |
| (For injections of 100 µL or greater, loop replacement) |
Reproducibility | CV 0.5 % or less (when 10 µL or more is injected) |
Number of samples | 100 |
Safety features | Air detection in wash solution and sample solution |
| Needle lock when the sample table is not loaded |
| Monitoring of 6-way and 4-way valve rotations |
Column Oven |
Method | Hot-air agitation (double structure), PID control (cascade control) |
Range | Room temperature +10 to 60 ºC (1 ºC steps) |
Accuracy | ±0.5 ºC (inside internal box) |
Precision | ±0.02 ºC (inside internal box) |
No. of columns accommodated | 7.8 mm I.D. × 30 cm column × 8 (Tosoh product) |
Safety features | Temperature sensor for oven temperature monitor. |
| Temperature control stops if abnormalities are detected |
| Gas sensor for stopping temperature control |
| Thermal fuse (70 ºC) for temperature control circuit block |
| Door status monitor: Temperature control stops when door is opened |
| Pan for liquid leakage |
Detector: RI detector |
Method | Brice-type, double-path, double-flow |
Range | 1.00 - 1.80 |
Flow cell | Material: Quartz glass |
| Volume: 2.5 µL |
| Maximum pressure: 0.5 MPa |
Noise | 2 × 10-9 RIU (THF sealed, response: 3.0 s) |
Drift | 1 × 10-7 RIU/h (THF, 1.0 mL/min during liquid supply)) |
Temperature control | OFF, 35, 40, 45 ºC |
Safety features | Leak sensor |
| Thermal fuse for circuit block |
Dimensions | 680 (W) × 550 (H) × 500 (D) mm (excluding projections) |
Weight | 95 kg |
Power | AC 100 – 240 V, 50/60 Hz, 500 VA |